User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: TA11 Overlab Gain Sub DATE: 16-Jun-97 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 3 NUMBER OF LINES: 34 CONFIGURATION: Fluke 5700A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 MATH mem1 = m[1] 1.002 DISP Connect the 5700 HI/LO OUTPUT to UUT channel [MEM1]. 1.003 IEEE SS5; 1.004 IEEE SP; 1.005 MESS Adjust the stimulus until the trace has deflected 120mm 1.006 MESS (24 major divisions) to the left on the printout. 1.007 5700 300.00mV N 2W 1.008 IEEE HLT; 1.009 MATH mem1 = 300 1.010 MEME 1.011 MEMC mV 1.5% 2.001 MATH mem1 = m[1] 2.002 IEEE #[MEM1]:ATTN=100 2.003 IEEE #[MEM1]:S=20; 2.004 MESS Adjust the stimulus until the trace has deflected 125mm 2.005 MESS (25 major divisions) to the left on the printout. 2.006 5700 250.00V N 2W 2.007 IEEE HLT; 2.008 MATH mem1 = 250 2.009 MEME 2.010 MEMC mV 1.5% 3.001 END